The fast solvent extraction instrument uses high temperature and pressure solvent extraction technology to quickly extract organic target components in solid or semi-solid samples with a solid-liquid extraction method. Because its extraction efficiency and sample throughput are far superior to other traditional extraction methods, it is a technology widely used in the market.
Extraction |
Sample handling capacity(g) |
Solvent consumption(ml) |
Solvent/sample ratio(ml/g) |
Average extraction time |
Soxhlet Extraction |
10-30 |
300-500 |
16-30 |
4-48h |
ULtrasonic extraction |
30 |
300-400 |
10-13 |
0.5-1h |
Microwave extraction |
5 |
30 |
6 |
0.5-1h |
Automatic Soxhlet extraction |
10 |
50 |
5 |
1-4h |
FSE-6 |
10-100 |
15-150 |
1.5 |
15min |
technical parameter:
1. 26 extraction positions (including 24 extraction pool positions and 2 cleaning positions);
2. The extraction tanks of 1, 5, 10, 22, 34, 66, 100ml specifications are optional;
3. Brand new T-type extraction tank: simple structure, easy to disassemble and assemble;
4. Collection tray: 26-position 250ml collection position, or 26-position 60ml collection position (adapter required), or custom combination of 26-position 250ml and 60ml collection position (adapter required);
5. Working temperature range: room temperature~210℃; heating furnace temperature control accuracy≤±1℃;
6. Working pressure range: 0~20Mpa;
7. Heating furnace body: adopts a brand-new 360-degree full-circle heating method, which can be quickly heated to the set temperature, while ensuring that the entire extraction tank is heated evenly, and has an over-temperature protection function;
8. High-pressure infusion pump: flow rate range 0-100ml/min, output pressure: normal pressure ~ 25Mpa;
9. Pressure control unit: built-in pressure control unit, overpressure protection system and gas-liquid isolation protection device;
10. Solvent controller: quaternary low pressure gradient, automatic switching of up to 4 kinds of solvents and solvent ratio and mixing of different proportions;
11. Liquid pipeline: It is treated by chemical passivation technology, and a certain proportion of non-strong acid extractant such as acetic acid and phosphoric acid can be used;
12. Control interface: touch control interface, easy to operate;
13. Collection bottle detection function: to ensure the safe collection of extracted samples;
14. The lighting function of the extraction position and the collection position area, which is convenient for the operator to understand the operation of the instrument in time;
15. The dust-proof design of the extraction tray and the collection tray protects the instrument from dust pollution when it is running or when it is not in use.